Combined Mechanical Characterizations Increases Sensitivity in the Assessment of Human Cartilage Degeneration

  • par garon@biomomentum.com

Sim S; Hadjab I; Chevrolat L-A; Masse M; Tong AL; Lavigne P; Garon M; Quenneville E; Buschmann MD
Accepted for a podium presentation at ORS 2017

Les titres, résumés et légendes des publications sont présentés en anglais.

Introduction: We published a recent study showing superior sensitivity of electromechanical and indentation (instantaneous response) assessments versus well-established techniques, including histological Mankin score, to characterize cartilage degeneration. This study aims to determine whether the combination of instantaneous, relaxation and equilibrium mechanical properties and friction measurements (surface integrity) could increase sensitivity to detect cartilage degeneration.
Methods: To ensure the presence of different degeneration grades, healthy and degenerated cartilage were isolated from human cadaveric (RTI Surgical Inc., FL, USA) and TKR samples, respectively. Four patellas were used because of their high homogeneity throughout their entire surface and importance in joint health. On the entire articular surface, an electromechanical QP mapping was first performed using the Arthro-BST. The surface was then graded by an orthopaedic surgeon in accordance with the ICRS grading system to assess cartilage degeneration. Subsequently, an automated indentation mapping was performed at matching positions. Ten osteochondral cores per ICRS grade (0 to 3; 4 being mostly bone was excluded from this study) were then harvested from the patellas. Each core was tested following the sequence previously developed where 12 parameters were extracted (Fig. 1). These parameters have been classified as shown in Table 1. A multiaxial mechanical tester, Mach-1 v500css (Biomomentum Inc.), was used for all mechanical tests. All statistical tests were done in SAS version 9.3 (SAS Institute Inc., NC, USA).
Results: A general linear model revealed that at least one parameter per category provide high to moderate significant prediction of cartilage degeneration through ICRS grades (Table 1). Moreover, the combination which includes a parameter from the instantaneous response and 1 or 2 parameters from the other categories improved model fit (R2), suggesting increased sensitivity in the differentiation of ICRS grades (Table 2).
Discussion: This study established that a combination of different mechanical responses, where one should be an instantaneous response, predicts cartilage degeneration better than a single parameter. These results may reflect the fact that the patellofemoral joint is subjected to both load-bearing and friction, corresponding to different modes of cartilage function.
Significance: There is a lack of gold standard techniques allowing a sensitive distinction of cartilage degeneration stages. This study on the combination of instantaneous response with relaxation, equilibrium response and/or surface integrity parameters could lead to a novel approach of assessing cartilage integrity in studies of cartilage repair or degeneration.

Utilisation des équipements

Arthro-BST a cartographié le paramètre quantitatif électromécanique (QP) sur les surfaces cartilagineuses de rotules humaines. Le Mach-1 Mechanical Tester (v500css) a réalisé une cartographie automatisée par indentation aux positions correspondantes, puis des essais mécaniques sur les carottes ostéochondrales prélevées. L’évaluation associait réponses mécaniques instantanées, de relaxation et à l’équilibre et mesures de frottement ; l’extrait ne précise pas d’adaptation de la trajectoire d’indentation à la normale locale.

Article et documents

Matériaux (2)

Configurations de test (8)

Produits (1)

Découvrez la recherche en pratique

Découvrez les études de cas

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.