Aller au contenu

An interfacial stress sensor for biomechanical applications

  • par garon@biomomentum.com

K Sundara-Rajan; A Bestick; G I Rowe; G K Klute; W R Ledoux; H C Wang; A V Mamishev
Measurement Science and Technology. 2012.

https://doi.org/10.1088/0957-0233/23/8/085701

25 citations · 0 en 2025–2026

Les titres, résumés et légendes des publications sont présentés en anglais.

This paper presents a capacitive sensor that measures interfacial forces in prostheses and is promising for other biomedical applications. These sensors can be integrated into prosthetic devices to measure both normal and shear stress simultaneously, allowing for the study of prosthetic limb fit, and ultimately for the ability to better adapt prosthetics to individual users. A sensing cell with a 1.0 cm 2 spatial resolution and a measurement range of 0–220 kPa of shear and 0–2 MPa of pressure was constructed. The cell was load tested and found to be capable of isolating the applied shear and pressure forces. This paper discusses the construction of the prototype, the mechanical and electrode design, fabrication and characterization. The work presented is aimed at creating a class of adaptive prosthetic interfaces using a capacitive sensor.

Utilisation des équipements

Un Mach-1 avec cellule Nano17 à six axes caractérisait un capteur d’interface capacitif à piliers de PDMS et électrodes sur circuits imprimés. Les chargements normaux et latéraux étalonnaient ses réponses à la pression et au cisaillement.

Matériaux (1)

Configurations de test (2)

Découvrez la recherche en pratique

Découvrez les études de cas

Report a problem with this page