Skip to content

An interfacial stress sensor for biomechanical applications

  • by garon@biomomentum.com

K Sundara-Rajan; A Bestick; G I Rowe; G K Klute; W R Ledoux; H C Wang; A V Mamishev
Measurement Science and Technology. 2012.

https://doi.org/10.1088/0957-0233/23/8/085701

25 citations · 0 in 2025–2026

This paper presents a capacitive sensor that measures interfacial forces in prostheses and is promising for other biomedical applications. These sensors can be integrated into prosthetic devices to measure both normal and shear stress simultaneously, allowing for the study of prosthetic limb fit, and ultimately for the ability to better adapt prosthetics to individual users. A sensing cell with a 1.0 cm 2 spatial resolution and a measurement range of 0–220 kPa of shear and 0–2 MPa of pressure was constructed. The cell was load tested and found to be capable of isolating the applied shear and pressure forces. This paper discusses the construction of the prototype, the mechanical and electrode design, fabrication and characterization. The work presented is aimed at creating a class of adaptive prosthetic interfaces using a capacitive sensor.

Equipment used

A Mach-1 with a six-axis Nano17 load cell characterized a capacitive interface sensor with PDMS pillars and printed-circuit electrodes. Combined normal and lateral loading calibrated its pressure and shear responses.

Materials (1)

Test configurations (2)

See Research in Action

Explore case studies

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.