Abstract

Objective: Intralesional steroid injections for vocal fold (VF) scarring are rapidly cleared, limiting their effectiveness. This study evaluates the efficacy of a light-activatable dexamethasone implant compared to a single steroid injection for VF scar healing.

Methods:

Ten rabbits underwent endoscopic VF injury and were treated with either a dexamethasone injection or a light-activatable implant; injured-only and no-injury VFs served as controls. On days 0 and 21, VF implants were irradiated with a near-infrared pulsed laser for 1 min. Larynges were harvested after 42 days. High-speed video (> 10 kHz) of excised larynges captured VF vibrations for kymographic analysis. Normal force, structural stiffness, and displacement were measured. Data were analyzed with appropriate statistical tests.

Results:

Compared to injury-only VFs, implant-treated VFs demonstrated significantly reduced normal force and stiffness (5.28 ± 0.77 mN vs. 2.34 ± 0.66 mN, p < 0.0001; 17.74 ± 2.45 mN/mm vs. 6.80 ± 1.32 mN/mm, p < 0.0001), and increased displacement at 1.96 mN (0.14 ± 0.02 mm vs. 0.29 ± 0.05 mm, p < 0.0001) along the injury zone. Implant-treated VFs exhibited larger amplitude ratios and improved tissue architecture than untreated scars. No differences in quantitative measures were observed between implant-treated and injection-treated groups.

Conclusion:

A light-activatable dexamethasone implant improved VF biomechanics, vibratory behavior, and histological outcomes in a rabbit injury model comparable to a single steroid injection. This novel approach shows promise for delivering controlled, repeatable therapy to VF scars.

Keywords:

Controlled‐release; drug delivery; therapeutic implant; vocal fold scarring.

Utilisation des équipements

Le Mach-1 Mechanical Tester a réalisé des indentations sphériques sur des plis vocaux de lapin excisés afin d’évaluer leur mécanique après traitement des cicatrices. Un capteur de charge uniaxial enregistrait les réponses sur une grille couvrant les régions glottique et sous-glottique. Les paramètres mesurés étaient la force normale, la rigidité structurelle et le déplacement; les échantillons étaient fixés dans des moules en résine sur mesure.

Article et documents

Matériaux (1)

Configurations de test (2)

Découvrez la recherche en pratique