Zinan He; Rosaire Mongrain; Simon Lessard; Boris Chayer; Guy Cloutier; Gilles Soulez
Medical Engineering and Physics 77 (2020), 60–68. Original publisher PDF hosted by the author laboratory (LBUM, CRCHUM). Electronic publication date 15 January 2020 confirmed by PubMed 31954613. Original abstract: English. Human detail applies to the surgical AAA tissue, not the synthetic hydrogel or silicone. Tensile tests were performed using a Bose Enduratec ELF 3200, not Mach-1.
https://doi.org/10.1016/j.medengphy.2019.12.005
4 citations · 2 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abdominal aortic aneurysm (AAA) is an asymptomatic condition due to the dilation of abdominal aorta along with progressive wall degeneration, where rupture of AAA is life-threatening. Failures of AAA endovascular repair (EVAR) reflect our inadequate knowledge about the complex interaction between the aortic wall and medical devices. In this regard, we are presenting a hydrogel-based anthropomorphic mockup (AMM) to better understand the biomechanical constraints during EVAR. By adjusting the cryogenic treatments, we tailored the hydrogel to mimic the mechanical behavior of human AAA wall, thrombus and abdominal fat. A specific molding sequence and a pressurizing system were designed to reproduce the geometrical and diseased characteristics of AAA. A mechanically, anatomically and pathologically realistic AMM for AAA was developed for the first time, EVAR experiments were then performed with and without the surrounding fat. Substantial displacements of the aortic centerlines and vessel expansion were observed in the case without surrounding fat, revealing an essential framework created by the surrounding fat to account for the interactions with medical devices. In conclusion, the importance to consider surrounding tissue for the global deformation of AAA during EVAR was highlighted. Furthermore, potential use of this AMM for medical training was also suggested.
Utilisation des équipements
Un Mach-1 v500csst mesurait la friction entre une gaine de cathéter hydrophile et un cryogel de PVA, trois silicones ou du tissu aortique anévrismal humain prélevé chirurgicalement. Les essais dans l’eau à température ambiante utilisaient 0,1–3 N et 0,5–50 mm/s sur 80 mm. La traction utilisait un testeur Bose distinct.
Matériaux (4)

Hydrogel

Vasculaire (veine and artère)

Caoutchouc et élastomère