Xingyu Hou; Jiaqi Zhu; Chengfeng Pan; Xin Wang; Zhenzhou Nie; Qianli Ma; Zifeng Zhang; Junnan Xue; Huihui Du; Kai Fung Chan; Chuan Fei Guo; Li Zhang
Sci Adv . 2026 Sep 11;12(37):eaef6473. . Epub 2026 Sep 9.
https://doi.org/10.1126/sciadv.aef6473
0 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract
Iontronic pressure sensors featuring high sensitivity are promising across different emerging fields. In applications requiring high-fidelity sensing such as surgeries, reliable pressure monitoring necessitates sensitivity and stability at the same time to ensure safety. However, existing iontronic sensors often face a trade-off between them-soft ionic materials with high conductivity often present pronounced viscoelasticity, resulting in a low sensitivity-to-drift-rate (S/D) ratio. Here, we report a polyelectrolyte-based iontronic pressure sensor that achieves a nearly one order-of-magnitude improvement in S/D ratio over state-of-the-art designs using a polarity-balance strategy in the polyelectrolyte to simultaneously regulate ionic conductivity and viscoelasticity. The resulting sensor exhibits ultrahigh sensitivity (119.5 per kilopascal) and exceptional stability (drift rate < 1%) under prolonged pressures up to 450 kilopascals, enabling nearly drift-free pressure detection for high-precision surgical operations in both surgeon-operated and robotic-assisted procedures. This work provides an extensible material design strategy to enhance sensitivity while suppressing signal drift in soft pressure sensors, with broad relevance to robotics, medical devices, and wearables.
Utilisation des équipements
Le Mach-1 Mechanical Tester a caractérisé des éprouvettes de polyélectrolyte par traction jusqu’à rupture, étirement–retour cyclique et relaxation des contraintes à déformation constante. Les mesures ont fourni le module de Young, la contrainte et la déformation à rupture, ainsi que le comportement d’hystérésis. Le système étant piloté en déplacement, la relaxation a servi à évaluer la résistance au fluage plutôt qu’un essai direct à contrainte constante.
Article et documents
Consulter dans PubMed
Matériaux (1)