Xingyu Hou; Jiaqi Zhu; Chengfeng Pan; Xin Wang; Zhenzhou Nie; Qianli Ma; Zifeng Zhang; Junnan Xue; Huihui Du; Kai Fung Chan; Chuan Fei Guo; Li Zhang
Sci Adv . 2026 Sep 11;12(37):eaef6473. . Epub 2026 Sep 9.

https://doi.org/10.1126/sciadv.aef6473

0 citations · 0 in 2025–2026

Abstract

Iontronic pressure sensors featuring high sensitivity are promising across different emerging fields. In applications requiring high-fidelity sensing such as surgeries, reliable pressure monitoring necessitates sensitivity and stability at the same time to ensure safety. However, existing iontronic sensors often face a trade-off between them-soft ionic materials with high conductivity often present pronounced viscoelasticity, resulting in a low sensitivity-to-drift-rate (S/D) ratio. Here, we report a polyelectrolyte-based iontronic pressure sensor that achieves a nearly one order-of-magnitude improvement in S/D ratio over state-of-the-art designs using a polarity-balance strategy in the polyelectrolyte to simultaneously regulate ionic conductivity and viscoelasticity. The resulting sensor exhibits ultrahigh sensitivity (119.5 per kilopascal) and exceptional stability (drift rate < 1%) under prolonged pressures up to 450 kilopascals, enabling nearly drift-free pressure detection for high-precision surgical operations in both surgeon-operated and robotic-assisted procedures. This work provides an extensible material design strategy to enhance sensitivity while suppressing signal drift in soft pressure sensors, with broad relevance to robotics, medical devices, and wearables.

Equipment used

The Mach-1 Mechanical Tester characterized polyelectrolyte specimens through tensile tests to fracture, cyclic stretching–recovery and fixed-strain stress relaxation. Measurements provided Young’s modulus, fracture stress and strain, and hysteresis behaviour. Because the system was displacement-controlled, stress relaxation was used to assess creep resistance rather than performing a direct constant-stress creep test.

Article and documents

Materials (1)

Test configurations (2)

See Research in Action

Explore case studies

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.