Analytical model for dental implant insertion torque

  • by garon@biomomentum.com

Baixuan Yang; Ainara Irastorza-Landa; Peter Heuberger; Heidi-Lynn Ploeg
Journal of the Mechanical Behavior of Biomedical Materials 131 (2022), 105223. Available online 14 April 2022. Publisher PDF deposited by the authors at Queen’s University. Mach-1 directly performed Exp-1 and two post-hoc extreme cases in Exp-2; other Exp-2 tests used Electroforce 3230-AT Series III.

https://doi.org/10.1016/j.jmbbm.2022.105223

12 citations · 5 in 2025–2026

Maximum insertion torque (IT) for threaded dental implants is a primary clinical measurement to assess implant anchorage, and strongly influences the clinical outcome. Insertion torque is influenced by surgical technique, implant designs, and patient factors such as bone density and quality. In this study, an analytical model was proposed for IT to estimate contributions from the thread and taper separately. The purpose of this study was to test if the analytical model could 1. differentiate the parallel-walled and tapered implant; and, 2. represent four factors: bone surrogate density, drill protocol, implant surface finish and cutting flute. The IT was modeled as the sum of the torques from the thread’s inclined plane and interface shear stress from the tapered body integrated over the surface area, respectively, with two main parameters: effective force, F′, F′ and effective pressure, p′. The effective force, relates to the clamping force from the thread, while the effective pressure, p′, associates with the contact pressure at the bone-implant interface. The model performed well (R² = 0.88–1.0) and differentiated between the parallel-walled (p′ = 0) and tapered implants (p′ = 0.12). The model’s parameters could individually represent the effects of the four factors. High bone surrogate density, two-step drill protocol, and rough surface increased both F′ and p′. The cutting flute had opposing effects on F′ and p′ (β₄ = 0.35 and −0.24, respectively); and therefore, had the lowest net effect on IT. The proposed analytical model therefore improves the understanding of the principal contributors to dental implant IT by considering thread and taper mechanics independently.

Equipment used

A Mach-1 V500CST measured dental-implant insertion torque in rigid polyurethane foam: parallel-walled and tapered implants at 12 rpm, plus two post-hoc high-torque cases at 7.5 rpm. Other factorial experiments used an Electroforce. Torsion denotes rotary insertion, not a pure torsional material test.

Materials (1)

Test configurations (1)

See Research in Action

Explore case studies

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.