Viscoelasticity and histology of the human cartilage in healthy and degenerated conditions of the knee

  • par garon@biomomentum.com

Michael Seidenstuecker; Julius Watrinet; Anke Bernstein; Norbert P. Suedkamp; Sergio H. Latorre; Anastasija Maks; Hermann O. Mayr
Journal of Orthopaedic Surgery and Research, November 2019, 14:256.

https://doi.org/10.1186/s13018-019-1308-5

37 citations · 10 en 2025–2026

Les titres, résumés et légendes des publications sont présentés en anglais.

Background: There are many studies on osteoarthritis, but only a few studies deal with human arthrosis, comparing the mechanical properties of healthy and diseased samples. In most of these studies, only isolated areas of the tibia are examined. There is currently only one study investigating the complete mapping of cartilage tissue but not the difference between instantaneous modulus (IM) in healthy and diseased samples. The aim of this study
is to investigate the relationship between the biomechanical and histological changes of articular cartilage in the pathogenesis of osteoarthritis.
Methods: The study compared 25 tibiae with medial gonarthrosis and 13 healthy controls. The IM was determined by automated indentation mapping using a Mach-1 V500css testing machine. A grid was projected over the sample and stored so that all measurements could be taken at the same positions (100 ± 29 positions across the tibiae). This grid was then used to perform the thickness measurement using the needle method. Samples were
then taken for histological examinations using a hollow milling machine. Then Giemsa and Safranin O staining were performed. In order to determine the degree of arthrosis according to histological criteria, the assessment was made with regard to Osteoarthritis Research Society International (OARSI) and AHO scores.
Results: A significant difference (p < 0.05) could be observed in the measured IM between the controls with 3.43 ± 0.36 MPa and the samples with 2.09 ± 0.18 MPa. In addition, there was a significant difference in IM in terms of meniscus-covered and meniscus-uncovered areas. The difference in cartilage thickness between 2.25 ± 0.11 mm controls and 2.0 ± 0.07 mm samples was highly significant with p < 0.001. With regard to the OARSI and AHO scores, the samples differed significantly from the controls. The OARSI and AHO scores showed a significant difference between meniscus-covered and meniscus-uncovered areas.
Conclusions: The controls showed significantly better viscoelastic behavior than the arthrotic samples in the measured IM. The measured biomechanical values showed a direct correlation between histological changes and altered biomechanics in gonarthrosis.
Keywords: Gonarthrosis, Articular cartilage, Biomechanical testing, Mapping, Indentation, Histology, OARSI score, AHO score

Utilisation des équipements

Le Mach-1 Mechanical Tester (V500css) a servi à cartographier par indentation automatisée le cartilage tibial humain de genoux arthrosiques et de témoins sains. L’indentation sphérique mesurait le module instantané sur la surface cartilagineuse. L’indenteur sphérique était ensuite remplacé par une aiguille pour mesurer l’épaisseur du cartilage aux mêmes points de la grille, en traversant le cartilage jusqu’à la limite de l’os sous-chondral.

Article et documents

Matériaux (2)

Configurations de test (2)

Produits (2)

Découvrez la recherche en pratique

Découvrez les études de cas

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.