M. Garon; A. Legare; E. Quenneville; T.J. Sims; A.P. Hollander; M.S. Shive; A. Restrepo; M.D. Buschmann
Transactions International Cartilage Repair Society, 5.3, September 2007, Warsaw, Poland.
https://doi.org/10.1016/s1063-4584(07)61267-9
0 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
This conference study examined the reliability and biological meaning of Arthro-BST measurements on one human cadaver knee. Five users performed repeated mappings at 102 positions. Twenty-seven cartilage disks were then tested mechanically and analysed for matrix composition and collagen crosslinks. Agreement between users was high, with an intraclass correlation of 0.87. Electrical measurements correlated with fibril stiffness, tissue thickness, collagen content and the ratio of mature to immature collagen crosslinks. The findings support reproducible assessment of site-specific cartilage function in this specimen.
Editorial synthesis based on archived ICRS conference abstract; not the authors’ verbatim abstract.
Utilisation des équipements
La présentation décrit l’enregistrement des potentiels d’écoulement par Arthro-BST lorsque son indenteur sphérique est comprimé manuellement contre le cartilage articulaire. Le capteur comporte 37 microélectrodes réparties sur la surface de l’indenteur. Ces extraits originaux documentent le principe de mesure, mais ne permettent pas d’établir le nombre d’échantillons, le protocole de cartographie répétée ou l’équipement utilisé pour les essais mécaniques distincts.