A. Changoor; J. P. Coutu; M. Garon; E. Quenneville; M. B. Hurtig; M. D. Buschmann
Journal of Biomechanical Engineering 133(6):061005
https://doi.org/10.1115/1.4004230
33 citations · 3 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
An equine explant model was used to compare electromechanical and mechanical detection of cartilage injury immediately after impact. Adult trochlear cartilage received low, medium or high impacts and was assessed with an arthroscopic streaming-potential device, compression testing and microscopy. Streaming-potential measurements decreased after medium and high impacts, whereas significant deterioration in fibril stiffness and permeability was mainly detected after high impacts. Electromechanical measurements also correlated with intrinsic mechanical properties and cartilage thickness. In this experimental setting, the probe detected impact-related changes more sensitively than compression tests on isolated cartilage disks.
Editorial synthesis based on the source text; not the authors’ verbatim abstract. Read the original source.
Utilisation des équipements
Arthro-BST a mesuré les potentiels d’écoulement du cartilage lors de l’indentation de surfaces trochléennes équines avant et après impact. Son embout hémisphérique comportait 37 microélectrodes en or, et deux utilisateurs ont répété les mesures aux mêmes sites. Le cartilage prélevé a ensuite été testé en compression non confinée avec le Mach-1 Mechanical Tester. Les impacts ont été appliqués séparément avec un impacteur sur mesure, et non avec ces instruments d’évaluation.
Article et documents
Consulter l’article
Matériaux (1)