Garon, M.; Cloutier, L.; Légaré, A.; Quenneville, E.; Shive, M. S.; Buschmann, M. D
ORS Transaction. 32: 629.
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
Five users repeatedly mapped cartilage from one human cadaver knee with Arthro-BST to assess operator dependence and relationships with mechanical properties. The instrument detected streaming potentials at ninety-two of the 102 mapped positions. An intraclass correlation of 0.87 indicated good agreement between users. Electrical measurements correlated with fibril stiffness and cartilage thickness, with the strongest relationship involving their ratio. Maps also reflected differences between joint surfaces and meniscal coverage. This conference report supports reproducible regional measurements in one specimen; broader clinical validation was still planned.
Editorial synthesis based on the source text; not the authors’ verbatim abstract. Read the original source.
Utilisation des équipements
Arthro-BST a enregistré les potentiels d’écoulement lors de l’indentation manuelle du cartilage d’un genou humain cadavérique. Une grille superposée à l’image d’une caméra guidait les cartographies répétées par cinq utilisateurs, et les signaux étaient intégrés pour obtenir l’intégrale du potentiel d’écoulement. Les disques de cartilage prélevés ont été testés avec le Mach-1 Mechanical Tester en compression non confinée, par rampes de relaxation, pour comparaison avec les mesures électromécaniques.
Matériaux (1)