Non-destructive electromechanical assessment (Arthro-BST) of human articular cartilage correlates with histological scores and biomechanical properties

  • par garon@biomomentum.com

S. Sim; A. Chevrier; M. Garon; E. Quenneville; A. Yaroshinsky; C.D. Hoemann; M.D. Buschmann
Osteoarthritis Cartilage, 22(11) 1926-35. (2014)

https://doi.org/10.1016/j.joca.2014.08.008

50 citations · 6 en 2025–2026

Les titres, résumés et légendes des publications sont présentés en anglais.

OBJECTIVE: The hand-held Arthro-BST™ device is used to map electromechanical properties of articular cartilage. The purpose of the study was to evaluate correlation of electromechanical properties with histological, biochemical and biomechanical properties of cartilage.

METHOD: Electromechanical properties (quantitative parameter (QP)) of eight human distal femurs were mapped manually ex vivo using the Arthro-BST (1 measure/site, 5 s/measure, 3209 sites). Osteochondral cores were then harvested from different areas on the femurs and assessed with the Mankin histological score. Prior to histoprocessing, cores were tested in unconfined compression. A subset of the cores was analyzed with polarized light microscopy (PLM) to assess collagen structure. Biochemical assays were done on additional cores to obtain water content and glycosaminoglycan (GAG) content. The QP corresponding to each core was calculated by averaging all QPs collected within 6 mm of the core center.

RESULTS: The electromechanical QP correlated strongly with both the Mankin score and the PLM score (r = 0.73, P < 0.0001 and r = -0.70, P < 0.0001 respectively) thus accurately reflecting tissue quality and collagen architecture. Electromechanical QP also correlated strongly with biomechanical properties including fibril modulus (r = -0.76, P < 0.0001), matrix modulus (r = -0.69, P < 0.0001), and log of permeability (r = 0.72, P < 0.0001). The QP correlated weakly with GAG per wet weight and with water content (r = -0.50, P < 0.0003 and r = 0.39, P < 0.006 respectively).

CONCLUSION: Non-destructive electromechanical QP measurements correlate strongly with histological scores and biomechanical parameters providing a rapid and reliable assessment of articular cartilage quality.

Utilisation des équipements

Arthro-BST a cartographié ex vivo les propriétés électromécaniques du cartilage fémoral distal humain en mesurant les potentiels d’écoulement induits par compression avec son indenteur sphérique muni de microélectrodes. Des carottes ostéochondrales ont ensuite subi des essais de relaxation des contraintes en compression non confinée sur le Mach-1 Mechanical Tester, leur partie osseuse étant fixée dans un support. Ces évaluations complémentaires permettaient des comparaisons avec les propriétés histologiques, biochimiques et biomécaniques.

Article et documents

Matériaux (2)

Configurations de test (1)

Produits (1)

Découvrez la recherche en pratique

Découvrez les études de cas

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.