David M. Kingsley; Caitlin H. McCleery; Christopher D.L. Johnson; Michael T.K. Bramson; Deniz Rende; Ryan J. Gilbert; David T. Corr
Journal of the Mechanical Behavior of Biomedical Materials, available online 10 January 2019
https://doi.org/10.1016/j.jmbbm.2019.01.003
33 citations · 15 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Cylindrical test specimens were punched from the prepared gels, using a 4-mm biopsy punch, to approximate the ratio of sample geometries (4-mm diameter, 2-mm thickness) specified by ASTM standard D575-91. Samples were characterized on a Mach-1 mechanical tester (Biomomentum Inc., Laval, QC, Canada) equipped with either a 70-N or 1.5-N load cell, depending on sample rigidity. Each test specimen was compressed to approximately 0.50 strain (absolute compression distance of 1.125 mm) at a constant rate of 0.03 mm/s. A stress-strain curve was constructed based on the measured force and displacement data (normalized to initial sample cross-sectional area and length, respectively). The elastic modulus was estimated from the linear region of this curve, which occurred between 0.05–0.25 strain for PDMS, and 0–0.10 strain for agarose. Six material test specimens were punched from three independently made gels, for all PDMS and agarose groups (n = 18 per group).
Indentation testing
Indentation characterization was performed on samples within 90-mm Petri dishes, using a Mach-1 mechanical tester (Biomomentum Inc., Laval, QC, Canada) with a spherical tip probe (6.35 mm diameter). A 1.5-N or 70-N load cell was used for all samples, depending on sample rigidity. To characterize, the probe height was first calibrated to a Petri dish containing no gel, to standardize the height of the probe. Gel tests began using the system's “Find Contact” mode, wherein the probe is lowered at a rate of 0.03 mm/s until a load was detected. Based on the determined initial sample height, the test was set to a total probe displacement of 30% strain, at a constant rate of 0.03 mm/s for all samples. The elastic moduli of indentation data were estimated using a Hertzian model, which accounts for the contact mechanics between a sphere and an elastic solid [12, 13]. To utilize the Hertz model, measured indenter force was plotted against the indentation depth raised to the 3/2 power, and the slope of the initial linear region, (between 0–0.1 strain for PDMS, and 0.03–0.09 strain for agarose), was used to solve for elastic modulus, E, in Eq. (2),
Utilisation des équipements
Le Mach-1 mesurait la réponse en compression d’échantillons de PDMS et de gels d’agarose.
Article et documents
Consulter l’article (PDF)
Matériaux (1)