Elizabeth Seed; Monica Mary Kerr; David Wheatley; Iain William Stewart; Sean McKee; Christopher McCormick
Biomedical Engineering, Mathematics And Statistics, Health and Wellbeing, Published: 9 Sept 2025
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract
1.0 INTRODUCTION
Heart valves are essential for blood circulation. Valvular heart disease (VHD) is a major global health issue, affecting over 40.5 million people. The primary treatment is surgical valve replacement, but current options have limitations. Mechanical heart valves (MHVs) can last over 40 years but require lifelong anticoagulation, while bioprosthetic heart valves (BHVs) avoid anticoagulation but typically become dysfunctional within 10-20 years [1].
The Wheatley Heart Valve (WHV) is a novel polymeric valve designed to combine MHV durability with BHV biocompatibility. Its unique S-shaped leaflet design enables manufacture from a single, flat polymer sheet, thereby making a range of materials and manufacturing methods possible. Mechanical and surface characterization of the candidate biomaterials are crucial steps towards future optimization of the WHV design and manufacture process.
This study has produced a series of Lubrizol Carbothane™ TPU flat sheets using different manufacturing methods. These sheets have then been subjected to tensile tests and surface contact angle (SCA) testing.
2.0 METHODS
2.1 Materials Carbothane™ TPU grades AC 4095A polymer pellets were obtained from The Lubrizol Corporation. The pellets were dissolved in N-N Dimethylacetamide (DMAC) at 20% and 30% (w/v). Flat sheets were then either 1) cast in glass petri dishes or 2) spin-coated on 4” silicon wafers, with the 30% solution first spun at 500 rpm for 40s and then 8000 rpm for 20s. All samples were left to cure for 24 hrs in a vacuum oven at 60°C. Finally, samples were laser cut to the desired shape prior to testing. For SCA testing, Polydimethylsiloxane (PDMS), banana leaves, and glass slides, were used as control samples.
2.2 Equipment For tensile testing, a Biomomentum Mach-1 Mechanical Tester was used, with a 0-250N load cell. Data was processed using Python (3.14), R (4.4.1), Microsoft Excel, GraphPad Prism (10.1.2), and Kinovea v2023.1.2.
2.3 Isotropy Tensile Testing 2 groups of 45 x 9 mm flat sheet cast AC 4095 A samples were cut (n=3) perpendicular to each other and arbitrarily assigned as the longitudinal and transverse axis. Samples were preconditioned (10 cycles: ramp speed 7.5 mm/s, pulled 1.5 mm) and then pulled 10 mm along the x-axis at 7.5 mm/s. The gauge length was 30 mm. Sample thicknesses were measured at 3 points per sample with a digital micrometer. Young’s Modulus (YM), E was calculated using equation 1. The thicknesses and YMs were each compared with a Welches t-test.
E= σ/(ε ), σ=F/A Where F is the force applied, A is cross-sectional area, σ is stress, ε is strain.
3.0 RESULTS
3.1 Isotropy Tensile Testing The mean thickness of the transverse and longitudinal axis groups (n=3) were 251 (SD = 7.0) µm and 230 (SD = 11.3) µm, respectively. There was no significant difference between the group thicknesses (p=0.0823, Welches t-test). Figure 1 shows the YM of each group; there was no significant difference between the longitudinal and transverse tests (61.87 MPa v 57.85 MPa, p>0.05).
3.2 Surface Contact Angle
The mean SCA of the tested sample groups are shown in Table 1.
4.0 DISCUSSION
The tensile testing showed that the flat-cast samples were isotropic. There were slight, insignificant, differences in YM between the transverse axis and the longitudinal axis. This difference may reflect the difference in thickness between the samples, introduced by variability in the manufacturing method. In future research, it may be beneficial to introduce anisotropy to better match the native valve.
SCA is an indicator of potential biocompatibility, which is useful for long-term, blood contacting and implantable heart valves. The SCA testing showed the AC 4095 A material to be hydrophobic, similar to the PDMS control. The spin coated AC 4095 A samples were more hydrophobic than the flat cast ones, consistent with Wenzel’s model, as the surface appeared smoother [2]. This study reproduced the Janus wettability results of banana leaves [3], providing a useful control for potential future studies of leaflet design research. Further investigation into droplet deformation could explore how material topography and isotropy are linked.
Future work will investigate the impact and potential benefits of different manufacturing methods and possible coatings on the mechanical behavior and haemocompatibility of the valve leaflets. The spin coated isotropy will be investigated. These data will be used in in-silico models of the valve to help inform design optimization [4].
5.0 CONCLUSIONS
The AC 4095A flat cast material is isotropic and AC 4095 A is hydrophobic. These data collected have provided a baseline which can be taken forward to define the optimum leaflet design in the Wheatley valve.
Keywords
Flexible biomaterials, Polymeric heart valve
Utilisation des équipements
Le Mach-1 Mechanical Tester a réalisé les essais de traction de feuilles de TPU Carbothane pour le développement de valves cardiaques polymériques avec une cellule de charge de 250 N.
Article et documents
Document
Consulter l’article
Matériaux (1)

Matériau
Plastique et polymère (feuille)
Configurations de test (1)
Configuration de test