Parisa R. Moshtagh; Behdad Pouran; Nicoline M. Korthagen; Amir A. Zadpoor; Harrie Weinans
Journal of Biomechanics 49(14) · September 2016
https://doi.org/10.1016/j.jbiomech.2016.09.020
39 citations · 4 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Mechanical properties of articular cartilage that are vital to its function are often determined by indentation tests, which can be performed at different scales. Cartilage tissue exhibits various types of structural, geometrical, and spatial variations that pose strict demands on indentation protocols. This study aims to define a reproducible micro-indentation protocol for measuring the effective (average) stiffness of the cartilage surface in a region around 1 mm2. We elucidated how different parameters such as indenter size, indenter depth, and the location of the indentation influence the effective elastic modulus measured in micrometer scale on rat knee cartilage. When an indentation was performed (50 μm radial probe, ≈10 μm indentation depth) at exactly the same location, the variability was less than 10%, even with a recovery period of 30 s. However, there was a high spatial variation and a small change of around 60 μm in location could change the modulus values up to as much as 10-20 fold. The effective elastic modulus of cartilage surface layer cannot therefore be reproducibly determined from a few indentations on a cartilage sample, and requires at least 144 (12×12) indentations for a soft spherical probe with a 50 μm radius. With higher depths, the spatial variation is slightly lower, allowing slightly lower number of indentations (≈80 measurements or a 9×9 frame) to provide a representative elastic modulus. Using this protocol, we determined an elastic modulus of 2.6±1.9 N/mm2 at the medial side versus a higher modulus of 4.2±2.6 N/mm2 at the lateral side of the tibia of 12 weeks old Wistar rats. Optimized indentation protocols similar to the one presented here are required for revealing such variations in the mechanical properties of cartilage with anatomical location.
Utilisation des équipements
L’étude examine l’influence de la taille de l’indenteur, de la profondeur d’indentation et du site de mesure sur le module élastique effectif du cartilage du genou de rat, avec un échantillonnage spatial répété pour caractériser la rigidité superficielle. Le résumé disponible n’identifie pas l’instrument utilisé ni le rôle du Mach-1 Mechanical Tester et ne précise aucune adaptation de la trajectoire à la normale locale.
Article et documents
Document
Consulter dans PubMed
Document
Consulter dans ResearchGate
Matériaux (2)

Matériau
Cartilage articulaire

Configurations de test (2)
Configuration de test
Cartographie 3D en indentation normale
Configuration de test