Wanqi Feng; Mark W. McDonald; Calum Anderson; David Smith; Robert L. Reuben; Duncan P. Hand; Yuhang Chen; William N. MacPherson
Proc. SPIE 13639, 29th International Conference on Optical Fiber Sensors, 136399I, Published: 22 May 2025
https://doi.org/10.1117/12.3061864
0 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract
Robot-assisted surgery (RAS) has emerged as a groundbreaking technique retaining much of the functionality of traditional open surgery in terms of visibility, flexibility and precision, whilst the minimally invasive incision adds benefits in healing time and reduced risk of complications. Force feedback plays a crucial role in ensuring surgical safety and effectiveness, as excessive or inappropriate forces can result in unwanted tissue damage, bleeding and/or poor outcome. In the work reported here, we embedded fibre Bragg grating (FBG) sensors along the length of a standard minimally invasive surgical gripper jaw to measure the strain resulting from a clamping force applied at the tip of the jaw. We have found that the FBG has sufficient sensitivity to monitor surgically relevant forces without excessive deformation. Future work will explore multi-axis strain measurements to monitor different types of force experienced by the gripper.
Utilisation des équipements
Des capteurs à réseaux de Bragg sur fibre optique ont été intégrés le long d’un mors de pince chirurgicale mini-invasive pour mesurer la déformation produite par une force de serrage appliquée à son extrémité. Le résumé disponible n’identifie pas l’équipement d’application de force ou d’essai de référence; un rôle d’essai ou d’étalonnage du Mach-1 Mechanical Tester ne peut donc pas être confirmé.