E. Quenneville; J.-S. Binette; M. Garon; A. Legare; M. Meunier; M.D. Buschmann
IEEE Trans Biomed Eng. 51(12): 2164-73.
https://doi.org/10.1109/tbme.2004.836522
14 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
This engineering paper describes the fabrication and characterization of a curved microelectrode array for measuring cartilage electrical responses during indentation. Platinum electrodes were incorporated into thin insulated circuits on spherical surfaces, allowing electrical recordings over the probe’s contact region. The researchers assessed electrical behaviour, including impedance and interactions between channels, and demonstrated cartilage measurements with the resulting array. The work addresses the manufacturing and instrumentation needed for an arthroscopic electromechanical probe. It is a device-development study and does not itself establish clinical diagnostic performance.
Editorial synthesis based on the source text; not the authors’ verbatim abstract. Read the original source.
Utilisation des équipements
Travaux historiques utilisant des réseaux de microélectrodes ou une première sonde arthroscopique pour étudier l’électromécanique du cartilage. Le dispositif décrit précède le Benchtop Arthro-BST actuel.
Article et documents
Consulter l’article (PDF)
Matériaux (1)