Denzel Ryan D. Cruz; Nour Awad; Nicholas Russell; Juliana El Sheikh; Logan Obermeyer; Peter J. Larson; Yoonjee C. Park; Gregory R. Dion
Journal Articles, 2026.
https://doi.org/10.1002/lary.70821
0 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Objective Innovative therapies are needed to improve the treatment of laryngotracheal stenosis. This study investigated the local tissue response of a novel injectable drug delivery platform in a leporine tracheal model. Methods Dye‐loaded polymer (empty) implants were delivered into the pretracheal tissue of three New Zealand White rabbits using an 18‐gauge needle and custom applicator to simulate peritracheal delivery adjacent to tracheal stenosis. Three additional rabbits underwent tracheal injury via electrocauterization to serve as comparative injury controls, and three additional rabbits were included as no‐implant, no‐injury tracheal controls. Animals underwent laryngoscopy and bronchoscopy on Days 0 and 14 to evaluate the primary outcome of local airway safety after implant placement. Secondary outcomes included ultrasound implant localization, indentation mapping of tracheal wall structural stiffness, and histological assessment. Results Implant‐treated tracheas ( n = 3) maintained lumen patency and organized tissue architecture without visible stenosis or microscopic inflammation, whereas injured tracheas ( n = 3) demonstrated luminal distortion and disorganized early remodeling. Implant location was confirmed by ultrasound. Indentation mapping showed lower mean anterior and posterior tracheal wall stiffness in implant‐treated specimens (5.28 and 4.47 mN/mm, respectively) compared with injured specimens (37.38 and 11.20 mN/mm, respectively). Posterior mucosal thickness was also lower in implant‐treated tracheas than injured tracheas (31.02 vs. 56.07 μm). Overall, the implants did not appreciably alter the native tracheal structure. Conclusion This novel injectable platform was well tolerated in the pretracheal space and preserved native tracheal structure, supporting future studies of therapeutic‐loaded implants for stenosis mitigation and treatment. Level of Evidence NA.
Utilisation des équipements
Un Mach-1 v500css équipé d’une cellule de charge uniaxiale de 1,5 N a réalisé une cartographie automatisée par indentation de sections trachéales antérieures et postérieures de lapin. Un indenteur sphérique de 0,5 mm avançait à 0,45 mm/s sur 0,1 mm. Les courbes force–déplacement quantifiaient la rigidité structurale locale de la paroi.
Matériaux (1)