Sim S; Chevrier A; Quenneville E; Garon M; Buschmann MD
P26, ICRS 2013, Turkey. (poster abstract 5677) (Best poster award)
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
This conference study compared Arthro-BST mapping with histological and mechanical characterization of human distal femoral cartilage. Eight donor specimens were mapped before osteochondral cores were harvested. Mankin grading was performed on 108 cores, and fifty underwent unconfined compression testing. Higher electromechanical QP values were associated with worse histological scores and lower fibril stiffness. The results connect nondestructive surface measurements with destructive reference assessments and support further evaluation of the probe for cartilage characterization. They do not establish diagnostic accuracy during routine patient arthroscopy.
Editorial synthesis based on archived ICRS conference abstract; not the authors’ verbatim abstract.
Utilisation des équipements
Arthro-BST a cartographié in vitro les propriétés électromécaniques fondées sur les potentiels d’écoulement du cartilage fémoral distal de huit donneurs humains. Un recalage par caméra reliait les valeurs locales du paramètre quantitatif (QP) aux carottes ostéochondrales prélevées ensuite, pour comparaison avec les scores de Mankin et les mesures en compression non confinée. Les valeurs proches du centre de chaque carotte étaient moyennées; l’instrument de compression n’est pas précisé.
Matériaux (1)