Garon, M
Institut de Génie Biomédical. Montréal, École Polytechnique de Montréal. Ph.D. Thesis
3 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
This doctoral thesis presents the design and validation of an arthroscopic probe that measures cartilage streaming potentials using thirty-seven microelectrodes. Development work included electrical safety and preparation for clinical evaluation. Measurements on excised equine and human cartilage were compared with mechanical and biochemical properties. A multiuser study on a human knee showed good agreement between operators, with an intraclass correlation of 0.87. Modelling also examined the effects of cartilage thickness, fibril stiffness and electrical behaviour outside the contact region. Together, these investigations support interpretation of the probe’s response as an indicator of cartilage function.
Editorial synthesis based on the source text; not the authors’ verbatim abstract. Read the original source.
Utilisation des équipements
Arthro-BST a servi à évaluer en laboratoire du cartilage excisé de genoux équins et humain. Des mesures en indentation ont caractérisé les distributions de potentiel électrique à l’extérieur du cartilage pour les comparer à une modélisation électromécanique. Une évaluation distincte de la répétabilité a fait intervenir cinq utilisateurs testant manuellement 102 sites sur cinq surfaces articulaires d’un genou humain, afin d’évaluer l’accord entre opérateurs.
Matériaux (1)