Comparison Between In Vitro and Simulated Arthroscopy Electromechanical Measurements of Human Articular Surfaces Using the Arthro-BST

  • par garon@biomomentum.com

Sim S; Becher C; Garon M; Quenneville E; Hurschler C; Buschmann MD
International Cartilage Repair Society 2016 in Sorrento, Italy. Poster 132

Les titres, résumés et légendes des publications sont présentés en anglais.

Purpose: The output of the Arthro-BST was originally streaming potentials integral (SPI) parameter (Abedian-2013). Since then, the output has changed to a quantitative parameter (QP) (Sim-2014). The purpose of this study was to reanalyzed old data to investigate whether QP values are reliable for assessing articular cartilage in a clinical context.
Materials & Methods: One closed cadaveric knee joint was obtained from Science Care (Arizona, USA). An orthopaedist opened the joint and defined 12 measurement sites on the distal femur with a permanent marker. The joint was then closed with stitches to perform arthroscopy. The distal femur was then isolated from the joint and in vitro measurements were performed at matched sites by 4 users (the orthopaedist and 3 engineers). Visual ICRS scoring and histological Mankin scores were obtained for all sites. In this study, we reanalyzed all electromechanical output from the Arthro-BST by taking the first valid measurement only. The Arthro-BST (Biomomentum Inc., Laval, Canada) calculates a quantitative parameter which is inversely proportional to the electromechanical activity of cartilage.
Results: An intraclass correlation coefficient (ICC) of 0.85 (95%CI: 0.68–0.95) was calculated among the four users, while an ICC of 0.89 (95%CI: 0.63–0.97) was found between QP values obtained under simulated arthroscopy and in vitro. Strong correlations were found between simulated arthroscopy and in vitro QP values (r=0.83, p=0.0009, N=12, Fig.1). Strong correlations were also obtained between the QP values obtained arthroscopically and the Mankin Score (r=0.84, p=0.0007, N=12, Fig.2a) and the ICRS Scoring (r=0.80, p=0.001, N=12, Fig.2b).
Conclusion: This study revealed strong correlations of QP values obtained in arthroscopy with histological Mankin and visual ICRS scores. In addition to correlation of QP with gold standards, high ICCs suggest that the Arthro-BST could be considered as a powerful tool for articular cartilage assessment.

Utilisation des équipements

Cette étude réanalysait des mesures existantes de l’Arthro-BST sur un genou cadavérique, plutôt que de présenter une nouvelle campagne d’acquisition. Les valeurs du paramètre quantitatif (QP) obtenues en arthroscopie simulée étaient comparées aux mesures in vitro effectuées aux mêmes sites du fémur distal par quatre utilisateurs. Seule la première mesure valide était retenue, et le QP était évalué par rapport aux scores visuels ICRS et histologiques de Mankin.

Article et documents

Matériaux (2)

Configurations de test (1)

Produits (1)

Découvrez la recherche en pratique

Découvrez les études de cas

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.