Characterization of the Vertical Stiffness Gradient in Cadaveric Human and Excised Canine Larynges

  • par garon@biomomentum.com

Jacob Michaud-Dorko; Gregory R. Dion; Charles Farbos de Luzan; Ephraim Gutmark; Liran Oren
Journal of Voice, 2024, , ISSN 0892-1997, Available online 7 September 2024

https://doi.org/10.1016/j.jvoice.2024.08.011

5 citations · 3 en 2025–2026

Les titres, résumés et légendes des publications sont présentés en anglais.

Abstract
Summary:
The elastic properties of the folds govern the characteristics of vocal fold vibrations. This study addresses existing gaps by investigating the Young’s modulus along the anterior-posterior direction in excised canine and cadaveric human vocal folds. Micro-indentation testing was conducted on six excised canines and three cadaveric human larynges. Multiple points along the medial glottal wall were indented to determine force-displacement, stress-strain relationships, and Young’s modulus as a function of Green’s strain. A vertical stiffness gradient was consistently observed in both canine and human samples, with higher stiffness in the inferior aspect compared with the superior aspect. The stiffness increased toward both the anterior and posterior directions from the mid-coronal plane, with a more pronounced increase at the posterior edge. Human vocal folds generally exhibited lower stiffness at low strains but were comparable to canine vocal folds at higher strains. These findings suggest that the canine larynx model is a reasonable representation of the human laryngeal tissues’ elastic property trends. This analysis of the vertical stiffness gradient in canine and human vocal folds provides valuable data for improving experimental and numerical models of phonation.
A Biomomentum Mach-1 (Biomomentum Inc., Laval, Quebec, Canada) mechanical tester with a 150-g single-axis load cell and a 7.5-mg resolution performed automated indentations on the nonplanar hemilaryngeal surfaces. For each vocal fold specimen (left and right), the vocal folds were indented 2 mm perpendicular to the surface using a spherical indenter with a radius of 2.5 mm at each specific data point. A normal indentation of 2 mm was chosen since typical glottal width for excised canines is between 3 mm and 4 mm for low and high subglottal pressure, respectively
Keywords:
Vocal fold elasticity; Young’s modulus; Anterior-posterior vertical stiffness gradient; Micro-indentation.

Utilisation des équipements

Le Mach-1 a cartographié la raideur locale de plis vocaux canins excisés et humains cadavériques. Le protocole déterminait l’orientation de la surface par quatre contacts voisins, puis adaptait l’indentation à la normale locale : une cartographie par indentation normale 3D.

Matériaux (1)

Configurations de test (1)

Découvrez la recherche en pratique

Découvrez les études de cas