M. Garon; A. Legare; E. Quenneville; T.J. Sims; A.P. Hollander; M.B. Hurtig; M.S. Shive; M.D. Buschmann
Transactions International Cartilage Repair Society, 22.6, September 2007, Warsaw, Poland.
https://doi.org/10.1016/s1063-4584(07)61329-6
0 citations · 0 en 2025–2026
Les titres, résumés et légendes des publications sont présentés en anglais.
Abstract — Editorial summary
This conference study compared Arthro-BST measurements with mechanical and biochemical properties at corresponding sites in two equine stifle joints. After electromechanical mapping, cartilage disks and rings were analysed for stiffness, permeability, glycosaminoglycans, collagen and collagen crosslinks. The instrument distinguished tibial plateau cartilage from femoral condyles and trochleae. Streaming potentials correlated with collagen quantity and mature crosslink content. Regional differences in mechanical properties and composition helped explain the electrical maps. The study provides laboratory evidence connecting probe measurements with cartilage function across joint surfaces.
Editorial synthesis based on archived ICRS conference abstract; not the authors’ verbatim abstract.
Utilisation des équipements
Arthro-BST a été utilisé manuellement in vitro pour mesurer les potentiels d’écoulement du cartilage des plateaux tibiaux, des condyles fémoraux et des trochlées de deux genoux équins. Des disques et anneaux de cartilage ont ensuite été prélevés aux sites mesurés pour comparaison avec des essais de relaxation en compression non confinée et des analyses biochimiques. L’extrait ne précise pas l’instrument utilisé pour les essais de compression.
Matériaux (1)