Measuring the coefficient of friction between two cartilage surfaces in rotation

Our articulations have evolved to permit a near frictionless movement of the joint; the coefficient of friction of cartilage is so low that it can be very difficult to measure. One of our clients has developed an innovative test protocol to measure both the static and the dynamic friction coefficients. This client was also interested in knowing the effect of the waiting time, before initiating the movement, on the static friction coefficient as well as the effect of the lubricant (PBS or synovial fluid) on both coefficients.

We have worked with him to adapt his test protocol and experimental setup to our Mach-1 v500ct multiaxial mechanical tester (see Image1). From the flat portion of the trochlear groove of bovine femoral heads, large cylindrical osteochondral cores (2 different diameters) were harvested (see Image2). The core with the large diameter was carefully press-fit into a clear flexible tubing (to create a wall to contain the lubricant) and fixed on a secured holder attached to the rotational stage of the Mach-1. The smaller diameter piece was attached under the multiaxial load cell using a custom attachment fixture. (See Image1)
After filling the chamber with lubricant (either PBS, a negative control, or synovial fluid), the upper cartilage surface was placed in contact with the opposing cartilage surface with a vertical offset load applied. Using the rotational stage, both surfaces were rotated against each other after predefined waiting periods. After some adjustments were made to consider the geometry of the test configuration, the coefficients of friction were calculated using the torque value along the vertical axis (Tz) divided by the vertical force (Fz) measured between the surfaces. The static coefficients of friction were calculated based on the required torque to initiate the movement after a certain resting period, while the dynamic coefficients were calculated using the torque measured when a specific rotation speed was reached.
The results obtained (Image3) were in line with the literature for the magnitude of the coefficient of friction and, as expected, confirms the positive impact of the synovial fluid in the reduction of the coefficients of friction in the articular joint.

Figures

Mach-1 apparatus for rotational friction testing between cartilage surfaces.
Figure 1 — Show caption

Mach-1 apparatus for rotational friction testing between cartilage surfaces.

Cartilage specimen preparation, lubricant bath and rotational friction test geometry.
Figure 2 — Show caption

Cartilage specimen preparation, lubricant bath and rotational friction test geometry.

Dynamic friction coefficient versus waiting time for cartilage specimens in PBS or synovial fluid.
Figure 3 — Show caption

Dynamic friction coefficient versus waiting time for cartilage specimens in PBS or synovial fluid.

Find Your Testing Solution

Explore preconfigured testers

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.