Skip to content

Indentation and thickness mapping on samples of different stiffnesses in a 96-well plate

The testing of biological materials as well as bacteriological, viral, and cell cultures is made easier with well plates. They contain many plastic-formed wells that are used as miniature test tubes. This is excellent for downsizing test space and for performing many tests at once. In this case study, a multiaxial mechanical tester model Mach-1 v500css was used to determine the thickness and stiffness of small samples of agarose gels in a 96-well plate. Multiple concentrations were tested from 0.8% to 1.2% and with volumes of these gels varying from 0.1 to 0.5 mL. The samples were made from 6 mm biopsy punched disks instead of pouring the Agar solution directly into the wells in order to ensure a flat surface for indentation testing.
The high accuracy of the mechanical tester was needed to obtain precise data on the mechanical properties of these samples with thicknesses smaller than 4 mm. The Mach-1 is a flexible system adapted for highly sensitive and precise testing of many types of samples (biomaterials, microspheres, pills, living tissues, etc.). The Scan function of its software, Mach-1 Motion, allows the automation of the indentation test in multiple wells consecutively.
In order to properly measure thicknesses and avoid errors caused by the misalignments of the well plate, the location (x, y) of the B1 well is determined and the height at the bottom of this and 2 other wells were found. These 3 (x, y, z) coordinates allow a plane to be fit under the plate which is used to determine the exact bottom z-coordinate of each well. Thus, thicknesses can be determined for each sample after testing. An indentation is performed using controlled amplitude and velocity in each well containing a sample. All this is considered in the development stage of the study. To perform the test, one can simply calibrate the Mach-1 at the reference well B1, and load a pre-saved sequence to perform the indentation consecutively in all the wells containing a sample.
By implementing a custom script on the obtained data, the thickness and indentation modulus of each sample was extracted using Hayes Elastic Model. The results obtained show that the modulus is higher for samples of higher concentration of agar. Further analyses can be performed by modifying the script or performing statistical analyses on the data.
If you are interested in performing multi-well plate scanning, please contact us. We can provided preprogrammed test sequences (Image 4) and Matlab code, Excel and examples results in a .zip file (Image 5).

Figures

Mach-1 setup for automated indentation of samples in a 96-well plate.
Figure 1 — Show caption

Mach-1 setup for automated indentation of samples in a 96-well plate.

Plate layout showing agar samples prepared at different concentrations and volumes.
Figure 2 — Show caption

Plate layout showing agar samples prepared at different concentrations and volumes.

Thickness and instantaneous modulus maps, with tabulated measurements for the multi-well samples.
Figure 3 — Show caption

Thickness and instantaneous modulus maps, with tabulated measurements for the multi-well samples.

Mach-1 software interface and test sequences for different multi-well plate formats.
Figure 4 — Show caption

Mach-1 software interface and test sequences for different multi-well plate formats.

Example folders, MATLAB scripts and position files supporting automated multi-well plate testing.
Figure 5 — Show caption

Example folders, MATLAB scripts and position files supporting automated multi-well plate testing.

Find Your Testing Solution

Explore preconfigured testers

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.