Evaluation of the Mechanical Properties of Human Hair

We were recently asked if our mechanical tester has the capability of evaluating the mechanical properties of single fiber human hair. In our lab, a scientist has been devoted to this feasibility study.
The hardest part of this work was to develop the proper gripping fixtures to easily mount the hair on the tester to be able to perform tensile and torsion testing. Following few attempts, it was found that the simplest and most reproducible mounting method was based on the use of a pair of off-the-shelf syringe needle tips. On the bench, the hair fiber was inserted through the two needles facing each other and spaced by a predefined gauge gap. A single drop of cyanoacrylate glue was applied in the plastic cavity of each needle to secure the hair in place. Two needle tip holders with the same tapered shape as a regular syringe were mounted on the Mach-1. These two holders were fastened facing each other at the center of a testing chamber under a 17N multiaxis load cell. The double-needle hair assembly was then firmly inserted over the holder (see Image 1).
After zeroing the load, and considering the relative position of the vertical stage, the initial length of the hair at full extension (when no load is applied) was measured using the Mach-1 Motion's "Find Contact" function. A sinusoidal deformation in the elastic region was applied for 30 cycles at 0.5, 1 and 2 Hz. A Dynamic Mechanical Testing analysis was conducted as per MA056-SOP04-D to extract the complex tensile storage and loss modulus components. Dynamic testing showed an increase in phase magnitude with increased frequency. This result was expected and explains an increase in plastic deformation (loss modulus) with increased frequency.
The tensile modulus and ultimate tensile strength were also measured and calculated (Mach-1 Analysis software) during a tensile load ramp to failure (see Image 2 - Tensile test output until failure). The results were found to be in accordance with literature (Lee, 2012): 930 MPa ultimate tensile strength, 6.5 GPa elastic modulus. Other tests were completed using a rotational stage to apply torsion to the hair until rupture occurred. For two samples, the results were 175 and 179 turns before failure. A camera feed with a 10X lens was used to visualize the torsion of the fiber in real time.
The plastic torsion of the hair remains even after breakage and can be visualized under a microscope (see Image 3 - Post-torsion photograph using 20x microscope).
During torsion, the generated torque around the Z-axis was too low to be measured by the 17N multiaxial load cell. However, the vertical contraction of the hair induced by torsion created a vertical contractile force that was measured by the Fz channel of the cell. It is assumed that through proper theoretical modelling of this test configuration, this axial load could be used to indirectly measure the torsional shear strength of the hair.
Different indirect ways of quantifying the torsional properties of fibres have also been proposed in literature. For example, by relating the initial gauge length to the number of rotations before rupture, a theoretical value for fracture angle can be obtained. From a basic analysis, the fracture angle was found to be between 26 and 33°.

Figures

Single hair mounted between grips for a tensile test.
Figure 1 — Show caption

Single hair mounted between grips for a tensile test.

Force-position curve recorded during tensile testing of a hair.
Figure 2 — Show caption

Force-position curve recorded during tensile testing of a hair.

Magnified view of a hair specimen with local variations in shape.
Figure 3 — Show caption

Magnified view of a hair specimen with local variations in shape.

Find Your Testing Solution

Explore preconfigured testers

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.