Characterization of the Mechanical Properties of Alginate Beads
Alginate beads are used extensively in the biomedical industry. Mechanical properties of alginate beads with diameters smaller than 1 mm can be precisely measured with the Mach-1.
Two types of beads with different a-L guluronic content or G-content (LVM with 13% and LVG with 70%) were tested under unconfined compression in 0.15 M NaCl and in 50 mM CaCl2 solutions following a 30 minutes equilibration. A compression ramp of 50% of the bead diameter was applied in 2 seconds and an equilibrium endpoint was assumed to be a load relaxation rate of 0.05 g/min.
The Mach-1 is designed for various testing protocols including stress relaxation, dynamic sinusoids and creep. A typical result of a stress relaxation test with an LVG sample in CaCl2 solution can be seen. Table 1 summarizes the mechanical properties of the two types of beads in the two testing solutions (refer to related resource file). N is the number of samples, D represents the sample diameter, Ft is the transient load, Feq is the equilibrium load, t1 is the time when half the peak load is obtained during the ramp and t2 is the time when half of the transient load is relaxed.
We have observed that LVG beads with 70% G-content were larger than LVM beads with 13% G-content. LVG beads were also stiffer than LVM beads in the two solutions. However, LVM beads seemed to have a weakening reaction with calcium. In fact, LVM beads are stiffer and relax more slowly in 0.15 M NaCl than in 50 mM CaCl2 . High G-content increases bead stiffness and appears to diminish sensitivity to the presence or absence of calcium.
The high accuracy of the mechanical tester was needed to obtain precise data on bead mechanical properties with diameters smaller than 1 mm. The Mach-1 is a flexible system adapted for highly sensitive and precise testing of many types of samples (biomaterials, microspheres, pills, living tissues, etc.) in controlled environments (incubators). Its ease of use and modular software offers unlimited testing possibilities and permits direct and rapid analyses of results.