Characterization of Electromechanical Properties of Connective Tissues - Part 2 - Meniscus

In this case study, the Mach-1 mechanical tester was used to determine the electromechanical properties of two connective tissues; the cartilage and the meniscus. Since tissue stiffness and compression-generated electric fields, or streaming potentials, are sensitive to the biochemical content of the tissue, the tester was coupled with an array of electrodes that can precisely measure these electric signals. The objective of this study was to establish links between biochemical composition and electromechanical properties (stiffness and streaming potentials) of normal and degenerated connective tissues. The second part of the study focuses on meniscus.
The meniscus is a tissue similar to cartilage but softer and less charged. In this case, the samples have been sectioned to five 3-mm disks (about 1-2 mm thick) from various specified regions. Dynamic stiffness under unconfined compression and the streaming potential distribution are measured. It should be noted that the streaming potential are around 10 times lower for meniscus than cartilage (refer to Part 1 of this case study). These results also indicate that the dynamic stiffness and the streaming potential profiles are higher on the cartilaginous surface of the inner part of the meniscus (disk 1 and 2) compared to middle zones (disks 3 and 5). It is expected that these results are directly related to the proteoglycan content of menisci.
The Mach-1 mechanical tester coupled with an array of electrodes can precisely determine the electromechanical properties of a large range of normal and degraded connective tissues.

Figures

Bovine meniscus specimens prepared for electromechanical characterization.
Figure 1 — Show caption

Bovine meniscus specimens prepared for electromechanical characterization.

Meniscus sections and locations of disc specimens used for testing.
Figure 2 — Show caption

Meniscus sections and locations of disc specimens used for testing.

Dynamic stiffness as a function of loading frequency for meniscus disc specimens.
Figure 3 — Show caption

Dynamic stiffness as a function of loading frequency for meniscus disc specimens.

Streaming potential amplitude versus radial position for the meniscus discs shown.
Figure 4 — Show caption

Streaming potential amplitude versus radial position for the meniscus discs shown.

Find Your Testing Solution

Explore preconfigured testers

Report a problem with this page

Your email is used only to follow up on this report.

Screenshots (optional)

Click this area and paste a screenshot, or drag files here. Up to 3 PNG, JPEG or WebP images; 5 MB per image. Hide confidential information first.

Please do not include passwords or customer information. Reports and screenshots are private.